Issue 4, 2011

Trace species detection in the near infrared using Fourier transform broadband cavity enhanced absorption spectroscopy: initial studies on potential breath analytes

Abstract

Cavity enhanced absorption measurements have been made of several species that absorb light between 1.5 and 1.7 µm using both a supercontinuum source and superluminescent light emitting diodes. A system based upon an optical enhancement cavity of relatively high finesse, consisting of mirrors of reflectivity ∼99.98%, and a Fourier transform spectrometer, is demonstrated. Spectra are recorded of isoprene, butadiene, acetone and methane, highlighting problems with spectral interference and unambiguous concentration determinations. Initial results are presented of acetone within a breath-like matrix indicating ppm precision at <∼10 ppm acetone levels. Instrument sensitivities are sufficiently enhanced to enable the detection of atmospheric levels of methane. Higher detection sensitivities are achieved using the supercontinuum source, with a minimum detectable absorption coefficient of ∼4 × 10−9 cm−1 reported within a 4 min acquisition time. Finally, two superluminescent light emitting diodes are coupled together to increase the wavelength coverage, and measurements are made simultaneously on acetylene, CO2, and butadiene. The absorption cross-sections for acetone and isoprene have been measured with an instrumental resolution of 4 cm−1 and are found to be 1.3 ± 0.1 × 10−21 cm2 at a wavelength of 1671.9 nm and 3.6 ± 0.2 × 10−21 cm2 at 1624.7 nm, respectively.

Graphical abstract: Trace species detection in the near infrared using Fourier transform broadband cavity enhanced absorption spectroscopy: initial studies on potential breath analytes

Additions and corrections

Article information

Article type
Paper
Submitted
01 Jul 2010
Accepted
18 Nov 2010
First published
13 Dec 2010

Analyst, 2011,136, 801-806

Trace species detection in the near infrared using Fourier transform broadband cavity enhanced absorption spectroscopy: initial studies on potential breath analytes

W. Denzer, G. Hancock, M. Islam, C. E. Langley, R. Peverall, G. A. D. Ritchie and D. Taylor, Analyst, 2011, 136, 801 DOI: 10.1039/C0AN00462F

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.