An evaporation-based microfluidic sample concentration method

Abstract

We present a method for sample concentration within microfluidic devices using evaporation-induced flow. Evaporation-induced flow is easy to incorporate into microfluidic designs and can be used to concentrate a wide variety of molecules. The practicality of this method was demonstrated with 0.2 μm fluorescent spheres and FITC-labeled BSA. Thirty two percent of the 0.6 μL fluorescent sphere suspension was concentrated into a well within a microfluidic device. In the same amount of time, 93% of the 0.6 μL FITC-labeled BSA solution was concentrated.

Article information

Article type
Communication
Submitted
11 Mar 2002
Accepted
15 Apr 2002
First published
30 Apr 2002

Lab Chip, 2002,2, 57-61

An evaporation-based microfluidic sample concentration method

G. M. Walker and D. J. Beebe, Lab Chip, 2002, 2, 57 DOI: 10.1039/B202473J

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.