Issue 8, 2002

Critical comparison between quadrupole and time-of-flight inductively coupled plasma mass spectrometers for isotope ratio measurements in elemental speciation

Abstract

An evaluation and critical comparison of the performance of a reference sequential mass analyzer (quadrupole) and a ICP-TOFMS instrument for ICP-MS multielemental analysis in different types of transient signals, mimicking those encountered in speciation analysis, is described. Several parameters including integration time, the number of isotopes “simultaneously” monitored, detection limits, precision and accuracy attainable for multiisotopic ratios measurements in transient signals are critically evaluated. Multielemental measurements in typical speciation analysis transient signals of different widths at half peak heights: 26 s (similar to those obtained with HPLC separations); 8 s (alike to those obtained in fast HPLC separations); and 3.5 s (mimicking those obtained in CE separations) have been performed. As expected, the analytical performance is bound to the concentration levels and number of isotopes to be measured. The best precisions were achieved using a TOF mass analyzer (≤ 1%) for more than 15 isotopes at concentration levels over 10 ppb to be measured in transient signals of less than 5 s. However, when fewer isotopes had to be measured (less than 7), even for very short transient signals (3.5 s of fwhm), the measured precision was always similar or even better using a quadrupole mass analyzer. The detection limits obtained (even for very short integration times) were always favourable to the quadrupole ICP-MS, while ICP-TOFMS demonstrated similar or better accuracy for isotope ratio measurements. Finally, the comparative speciation analysis of a mixture of organometallics of mercury, lead and tin by GC-ICP-MS (0.7s fwhm) is also investigated and applications to real speciation problems are discussed.

Article information

Article type
Paper
Submitted
19 Mar 2002
Accepted
05 Jun 2002
First published
25 Jun 2002

J. Anal. At. Spectrom., 2002,17, 950-957

Critical comparison between quadrupole and time-of-flight inductively coupled plasma mass spectrometers for isotope ratio measurements in elemental speciation

M. Vázquez Peláez, J. M. Costa-Fernández and A. Sanz-Medel, J. Anal. At. Spectrom., 2002, 17, 950 DOI: 10.1039/B202748H

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements