Comparison of interface cones for analysis of sodium-rich samples using quadrupole ICP-MS

Abstract

This study compares analytical figures of merit such as sensitivity, background, % 140Ce16O+/140Ce+, 40Ar16O+, time dependent and matrix induced interference effects for solutions containing up to 500 mg L−1Na for a conventional sampler cone designated as the High Performance Interface (HPI) with that obtained with a so-called Xi cone. HPI counts were higher than those observed using the Xi cone by factors of about 10 at low mass, the difference decreasing with increasing mass, and at high mass they were approximately equal to those observed with the Xi cone. As a consequence of low sensitivity when using the Xi interface, there is potential for determining high concentrations of low mass elements. The 4-times lower 40Ar16O+ signals and up to two times lower % 140Ce16O+ of the Xi cone are important analytical figures of merit for minimizing polyatomic ion interferences.

% 140Ce16O+/140Ce+ varied considerably with depth of sampling and were much higher in the HPI cone than in the Xi cone at large depths of sampling. Na-induced variations were mass dependent and larger in the Xi cone than in the HPI cone, amounting to about 25 and 60%, respectively. Long-term variations for 250 min amounted to about 10 and 25%, respectively. It is concluded that the HPI cone when coupled to a robust ICP (power > 1.3 kW, nebulizer gas flow rate < 0.85 L min−1) is more suited for routine analysis of environmental samples. However, in cases where oxide ion interferences must be reduced and high concentrations of low mass analytes are to be determined, the Xi cone has advantages.

Graphical abstract: Comparison of interface cones for analysis of sodium-rich samples using quadrupole ICP-MS

Article information

Article type
Technical Note
Submitted
24 Mar 2009
Accepted
23 Jul 2009
First published
02 Sep 2009

J. Anal. At. Spectrom., 2009,24, 1558-1563

Comparison of interface cones for analysis of sodium-rich samples using quadrupole ICP-MS

I. Brenner, J. Pacheco and M. Valiente, J. Anal. At. Spectrom., 2009, 24, 1558 DOI: 10.1039/B905919A

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements