A new analytical bias correction for in situ Sr isotope analysis of plagioclase crystals using laser-ablation multiple-collector inductively coupled plasma mass spectrometry

Abstract

A new analytical protocol was developed for correcting baseline-induced biases during the analysis of Sr isotope ratios in plagioclase using excimer laser-ablation Aridus-dual-intake-system multiple-collector inductively coupled plasma mass spectrometry (LA-MC-ICP-MS). Residual analytical biases of 84Sr/86Sr and 87Sr/86Sr were observed after applying on-peak background subtractions and mass-fractionation corrections using the conventional 86Sr/88Sr exponential internal normalization. The residual biases occurred only for samples analysed with LA and not for solution analyses using Aridus with the same instrumental setup. Based on observations from the ablation of NaCl and olivine crystals, we concluded that this was due to suppression and enhancement of the Kr baseline by loading of the LA sample aerosols and by the introduction of Kr from the samples, respectively. We also found that both the 84Sr/86Sr and 87Sr/86Sr isotope ratios were affected proportionally by the baseline biases of the LA analyses of an isotopically/compositionally homogeneous anorthite plagioclase (MkAn) from Miyakejima, Japan, and similar results were seen in theoretical calculations. Therefore, the bias correction for the target 87Sr/86Sr ratios was available using the shift in the simultaneously measured 84Sr/86Sr from the natural ratio. We then determined the correlation factors between 84Sr/86Sr and 87Sr/86Sr using numerical simulations and confirmed them by analysing MkAn, which closely reproduced the theoretical factors. By applying the new theoretical correction protocol, the accurate measurement of unknown plagioclase samples was possible without using any standard. This was confirmed by comparison with the results of micro-milling thermal ionization mass spectrometry (MM-TIMS) of the same plagioclase crystals with various compositions. The new correction technique improved reproducibility by a factor of four, providing a basis for an accurate Sr isotope analysis using LA-MC-ICP-MS.

Graphical abstract: A new analytical bias correction for in situ Sr isotope analysis of plagioclase crystals using laser-ablation multiple-collector inductively coupled plasma mass spectrometry

Supplementary files

Article information

Article type
Technical Note
Submitted
29 Oct 2012
Accepted
20 Mar 2013
First published
21 Mar 2013

J. Anal. At. Spectrom., 2013,28, 945-957

A new analytical bias correction for in situ Sr isotope analysis of plagioclase crystals using laser-ablation multiple-collector inductively coupled plasma mass spectrometry

J. Kimura, T. Takahashi and Q. Chang, J. Anal. At. Spectrom., 2013, 28, 945 DOI: 10.1039/C3JA30329B

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements