Direct determination of chromium in empty medicine capsules by tungsten coil atomic emission spectrometry

Abstract

Tungsten coil atomic emission spectrometry (WCAES) is used to determine trace levels of Cr in medicine capsules. No sample preparation other than 10 min vortex mixing with HNO3 and H2O2 is required. Matrix decomposition is carried out directly on the atomizer to minimize matrix effects and improve precision and accuracy. The entire analysis takes less than 12 min per sample including sample preparation. The instrumental setup is based on a tungsten filament extracted from 150 W, 15 V microscope light bulbs, solid state power supply, fused silica lens, crossed Czerny–Turner spectrograph and a thermoelectrically-cooled charge-coupled device detector. The limits of detection (LOD) and quantification (LOQ) for Cr are 3 and 9 µg L−1, which correspond to procedure LOD and LOQ of 0.4 and 1 µg g−1, respectively. These values allow for quantification at a level 28-fold lower than the maximum daily Cr intake from medicine and medicine formulations recommended by the United States Pharmacopeial Convention (USP). The procedure's linear dynamic range is between 3–100 µg L−1, with a R2 = 0.9992. It was applied to different empty medicine capsule samples and WCAES results were not significantly different from values found by inductively coupled plasma optical emission spectrometry (ICP OES) by applying a single-factor ANOVA test at the 95% confidence level (p > 0.05). The precision, calculated as %RSD, is between 5.2–7.4%. This is a simple, fast, efficient and potentially portable analytical procedure that may be used in quality control and regulation enforcement applications.

Graphical abstract: Direct determination of chromium in empty medicine capsules by tungsten coil atomic emission spectrometry

Supplementary files

Article information

Article type
Technical Note
Submitted
23 Mar 2015
Accepted
09 Apr 2015
First published
09 Apr 2015

J. Anal. At. Spectrom., 2015,30, 1395-1399

Author version available

Direct determination of chromium in empty medicine capsules by tungsten coil atomic emission spectrometry

D. A. Goncalves, J. Gu, M. C. dos Santos, B. T. Jones and G. L. Donati, J. Anal. At. Spectrom., 2015, 30, 1395 DOI: 10.1039/C5JA00100E

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements