Characterization of inductively coupled plasma time-of-flight mass spectrometry in combination with collision/reaction cell technology – insights from highly time-resolved measurements

Abstract

We present results from studies investigating the capabilities of inductively coupled plasma time-of-flight mass spectrometry (ICP-TOFMS) in combination with collision/reaction cell technology (CCT). Experiments were carried out using various sample introduction techniques including high- and low-dispersion laser ablation and microdroplet generation. Specifically, we investigated H2 as a reaction gas and He as a collision gas for their effects on reduction of background species and interferences, limits of detection (LODs), quantification capabilities, and structure of short transient signals. With H2 as reaction gas, argon-based ions were suppressed. Ar+ and Ar2+ signals could be attenuated to intensity levels in the hundreds and single-digit counts per second range, respectively. Selective suppression of Ar-based ions gives access to the most abundant isotopes of Ca and Se. Furthermore, the attenuation of the Ar+ signals allows the instrument to be operated without m/z-selective attenuation of the ion beam prior to TOF analysis, which enhances transmission of isotopes near m/z 40. It was also found that application of flow rates ≤4 mL min−1 of H2, He or mixtures of these gases results in collision-induced focusing, and enhances sensitivities by a factor of 1.5 to 2 and mass resolving power by up to 16%. Use of CCT enabled an improvement in the LOD for 40Ca of more than three orders of magnitude. The LOD for 80Se was improved by more than one order of magnitude. For many of the elements contained in NIST SRM 610, LODs were lower by a factor of two to four. However, for most elements, improvement in quantification accuracy was not observed. Experiments with microdroplet sample introduction demonstrated that the mass-dependent ion transit times between the ICP and the TOF extraction region are affected by the amount of He buffer gas in the collision cell. Changes in transit times, as well as signal broadening, were observed on a time scale of tens to hundreds of μs. The duration of individual aerosol plumes from low-dispersion laser ablation, however, remained practically unaffected from collisional effects and imaging at 100 Hz laser repetition rate with baseline-separated aerosol plumes is possible.

Graphical abstract: Characterization of inductively coupled plasma time-of-flight mass spectrometry in combination with collision/reaction cell technology – insights from highly time-resolved measurements

Supplementary files

Article information

Article type
Paper
Submitted
18 Aug 2018
Accepted
07 Nov 2018
First published
12 Nov 2018

J. Anal. At. Spectrom., 2019,34, 135-146

Characterization of inductively coupled plasma time-of-flight mass spectrometry in combination with collision/reaction cell technology – insights from highly time-resolved measurements

M. Burger, L. Hendriks, J. Kaeslin, A. Gundlach-Graham, B. Hattendorf and D. Günther, J. Anal. At. Spectrom., 2019, 34, 135 DOI: 10.1039/C8JA00275D

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements