Issue 3, 1991

Application of a high resolution inductively coupled plasma mass spectrometer to the measurement of long-lived radionuclides

Abstract

Some long-lived radionuclides, such as 99Tc, 226Ra, 232Th, 237Np, 238U, 239Pu and 240Pu were measured using high resolution inductively coupled plasma mass spectrometry (HR-ICP-MS). By using HR-ICP-MS with an ultrasonic nebulizer, the detection limits of these nuclides were 0.002–0.02 pg ml–1 and the sensitivities were 10 times better than those obtained using HR-ICP-MS without the ultrasonic nebulizer. More accurate isotopic data were also obtained using HR-ICP-MS than with quadrupole ICP-MS at lower concentrations of the analyte because of the improvement in counting statistics that can be obtained with HR-ICP-MS due to the greater efficiency of ion transmission. A comparison of the measurement of the 240Pu to 239Pu ratio is shown.

Article information

Article type
Paper

J. Anal. At. Spectrom., 1991,6, 205-209

Application of a high resolution inductively coupled plasma mass spectrometer to the measurement of long-lived radionuclides

C. Kim, R. Seki, S. Morita, S. Yamasaki, A. Tsumura, Y. Takaku, Y. Igarashi and M. Yamamoto, J. Anal. At. Spectrom., 1991, 6, 205 DOI: 10.1039/JA9910600205

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements