An adaptive breath sampler for use with human subjects with an impaired respiratory function

Abstract

An adaptive sampler for collecting 2.5 dm3 samples of exhaled air from human subjects with an impaired respiratory function is described. Pressure in the upper respiratory tract is continuously monitored and the data used to control an automated system to collect select portions of the expired breathing cycle onto a mixed bed Tenax and Carbotrapadsorbent trap for analysis by GC-MS. The sampling approach is intended for use in metabolomic profiling of volatiles in human breath at concentrations greater than µg m–3. The importance of experimental reproducibility in metabolomic data is emphasised and consequently a high purity air supply is used to maintain a stable exogenous volatile organic compound profile at concentrations in the range 5 to 30 µg m–3. The results of a 90 day stability study showed that exogenous VOCs were maintained at significantly lower levels (40 times lower for isopropyl alcohol) and with significantly higher reproducibility (80 times lower standard deviation for isopropyl alcohol) than would have been be the case if ambient air had been used. The sampling system was evaluated with healthy controls alongside subjects with chronic obstructive pulmonary disease. Subjects were able to breathe normally with control subjects observed to breathe at a rate of 9 to 17 breaths per minute, compared to 16 to 30 breaths per minute for subjects with COPD. This study presents, for the first time, observations and estimates of intra-subject breath sample reproducibility from human subjects. These reproducibility studies indicated that VOCs in exhaled breath exhibit a variety of dynamic behaviours, with some species recovered with a RSD <30%, while other species were observed to have significantly more variable concentrations, 30 to 130% RSD. The approach was also demonstrated to reliably differentiate the differences in the VOC profiles between alveolar and dead space air.

Graphical abstract: An adaptive breath sampler for use with human subjects with an impaired respiratory function

Article information

Article type
Paper
Submitted
16 Jun 2006
Accepted
05 Oct 2006
First published
06 Nov 2006

Analyst, 2007,132, 153-163

An adaptive breath sampler for use with human subjects with an impaired respiratory function

M. Basanta, T. Koimtzis, D. Singh, I. Wilson and C. L. P. Thomas, Analyst, 2007, 132, 153 DOI: 10.1039/B608608J

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements